The study summarized here evaluates post-staining analytical stability. Its findings inform research workflow discussions under the tested conditions; they do not establish diagnostic performance or patient outcomes.
Technical-stability study
Follow the same technical tests over time.
The study followed separately prepared technical tests from an early acquisition through later post-staining measurements.
Semen specimensBiorepository samples in the technical study
Technical testsIndependently pipetted preparations
Minute originReference acquisition for each test
Multiple technical tests came from the same specimens. The 70 preparations should not be read as 70 independent specimen donors.
The 60-minute observation
Read the mean alongside retention and discordance.
At 60 minutes, the mean DFI change was +1.04 percentage points relative to origin. Individual results varied: 81.4% were within ±5 percentage points, and 21.4% had a different DFI category.
Technical interpretationThe observations support a practical post-staining acquisition window of up to 60 minutes under the evaluated conditions. The group mean does not imply every result stayed within tolerance or retained its category.
All evaluated timepoints
Later measurements show greater DFI drift.
Both retention and category discordance are shown alongside the mean changes and concordance. The 120- and 240-minute observations were exploratory.
| Time after staining | DFI mean Δ vs origin | DFI Lin CCC | HDS mean Δ vs origin | HDS Lin CCC | DFI within ±5 pp | DFI category discordance |
|---|---|---|---|---|---|---|
| 15 min | +0.11 pp | 0.973 | −0.50 pp | 0.984 | 91.4% | 15.7% |
| 30 min | +0.28 pp | 0.959 | −0.68 pp | 0.981 | 87.1% | 17.1% |
| 60 min | +1.04 pp | 0.946 | −0.81 pp | 0.972 | 81.4% | 21.4% |
| 120 min | +3.40 pp | 0.797 | −1.05 pp | 0.948 | 76.7% | 30.0% |
| 240 min | +6.42 pp | 0.652 | −1.20 pp | 0.936 | 58.3% | 48.3% |
pp = percentage points. CCC = Lin’s concordance correlation coefficient. HDS retention within ±5 percentage points was 100% at every evaluated timepoint. DFI category discordance includes observations near category boundaries. The table summarizes the supplied study findings; it does not display confidence intervals or per-timepoint denominators.
Scope of the evidence
An analytical-stability study with defined limits.
What was evaluated
- Post-staining measurement stability
- Agreement with an early origin
- Retention within a practical tolerance
- Research category reproducibility over time
What was not established
- Clinical sensitivity or specificity
- A patient diagnosis
- Treatment selection or effectiveness
- Pregnancy or live-birth outcomes
Applying the findings
- Evaluated protocol and instrument conditions matter
- Technical replicates and specimens are distinct
- Mean agreement does not describe every result
- Each laboratory evaluates its research workflow
Study context
Review the methods behind the numbers.
This summary draws on the technical-stability study materials supplied to Varen. Contact us for the available methods and supporting technical information when assessing the workflow for your research.
FRAGSIGN and SDF-Index are For Research Use Only. Not for use in diagnostic procedures.
Request technical information
Ask about the evaluated workflow conditions, analysis methods, and available study materials.
Contact Varen